                  Photovoltaic Auto safety comprehensive tester

                  Product Description:Name:Photovoltaic Auto safety comprehensive testerType:BND-CS9933EG-15Standards: IEC61730; IEC61215; IEC 61646Description of PV safety tester:Photovoltaic safety tester is used to test PV module, cable, connector, terminal box, …etc which can deal with high voltage and high sensitivity insulation test. There are two functions that are high voltage test and insulation resistance test in total.
                  Description

                  Product Description:

                  Name:Photovoltaic Auto safety comprehensive tester

                  Type:BND-CS9933EG-15

                  Standards: IEC61730; IEC61215; IEC 61646


                  Description of PV safety tester:

                  Photovoltaic safety tester is used to test PV module, cable, connector, terminal box, …etc which can deal with high voltage and high sensitivity insulation test. There are two functions that are high voltage test and insulation resistance test in total.

                  According to the IEC61730 standard(photovoltaic module safety qualification—part I: Requirements for construction which describes the fundamental construction requirements for photovoltaic modules in order to provide safe electrical and mechanical operation during their expected lifetime), the definition of PV module application is divided into A, B, C.

                  The current measuring terminal can choose the grounding mode or the floating mode. Insulation voltage withstand test of photovoltaic module is based on the application of IEC61215/61646 and the maximum voltage of the system.

                  Maximum test voltage for the application class A:2000V plus 4 times the maximum system voltage

                  For the maximum voltage application category B:1000V system 2 times

                  According to the calculation of a maximum voltage of 8000V, CS9923GS, CS9923G is based on this standard production

                  According to the calculation of the maximum voltage B 6000V, CS9922G, CS9922GS is based on this standard production


                  Performance characteristics:

                  ◆With safety anti-electric wall function, it can automatically detect whether the input power is correct

                  ◆For photovoltaic system panels, controllers, and battery packs

                  ◆After the DC withstand voltage and insulation resistance tests are completed, the tester can quickly discharge within 0.2 seconds to protect the safety of the operator

                  ◆With current lower limit alarm function, it can prevent the open circuit of the test line from causing misjudgment of the tested components during the test. The output voltage has voltage upper limit and voltage lower limit alarm functions

                  ◆With keyboard lock function to prevent the operator from modifying the test parameters and ensure that the tested object is tested according to the set parameters

                  ◆Linear power amplifier drives the output of sinusoidal voltage (current), and the crest factor is within the range of 1.3 to 1.5

                  ◆Adopts 240*64 graphic LCD display and humanized interface design

                  ◆With Chinese and English display interfaces, it can meet the different needs of different users.

                  ◆Test parameters can be saved in advance: 30 files can be set, and each file can have 99 test steps

                  ◆Optional RS232, RS485, USB interface, standard PLC interface

                  pv-auto-safety-tester (1).jpg


                  The Specification of the PV safety test system:

                  Model

                  BND-CS9933EG-15









                  ACW


                  The output voltage

                  Range

                  0.050kV~6.500kV

                  Accuracy

                  ±(1.5% reading + 5V)

                  Resolution

                  1V

                  Maximum output power

                  130VA(6.500kV/20mA)

                  Maximum rated current

                  20mA

                  Lower current range

                  0~20mA, 0=no lower limit

                  Current range

                  200uA, 2mA, 20mA

                  Output waveform

                  Sine wave

                  Output waveform distortion

                  ≤2% (no load or pure resistive load)

                  Crest factor

                  1.3~1.5

                  Output signal type

                  DDS+ amplifier

                  Voltage rise time

                  0.3s~999.9s 0=voltage rise time off

                  Test time

                  0.3s~999.9s 0=continuous test

                  Voltage fall time

                  0.3s~999.9s 0=voltage fall time off

                  Interval time

                  0.0s~999.9s 0=interval time off

                  Output voltage mode

                  N-mode, G-mode







                  DCW






                  The output voltage

                  Range

                  0.050kV~6.500kV

                  Accuracy

                  ±(1.5%Value +5V)

                  Resolution

                  1V

                  Maximum output power

                  65W(6.500kV/10mA)

                  Maximum rated current

                  10mA

                  Current range

                  2uA、20uA、200uA、2mA、10mA

                  Ripple factor

                  ≤5%

                  Discharge time

                  ≤200ms

                  Maximum charge current

                  10mA

                  Voltage rise time

                  0.3s~999.9s 0=voltage rise time off

                  Test time

                  0.3s~999.9s 0=continuous test

                  Voltage drop time

                  0.3s~999.9s 0=voltage drop time off

                  Interval time

                  0.0s~999.9s 0=interval time off

                  Delay alarm time

                  0.3s~999.9s 0=delay alarm time off

                  Output voltage mode

                  N mode, G mode


                  IR



                  The output voltage

                  Range

                  0.050kV~1.500kV

                  Accuracy

                  ±(1.5% reading +5V)

                  Resolution

                  1V

                  Maximum upper limit setting value

                  50.00GΩ




                  Maximum and minimum setting values

                  50.00GΩ

                  Minimum setting values

                  1MΩ

                  Voltage rise time

                  0.3~999.9s 0=voltage rise time off

                  Test time

                  0.3s~999.9s 0=continuous test

                  Interval time

                  0.0s~999.9s 0=interval time off

                  Automatic gear switching

                  Can be set to on or off

                  Discharge time

                  ≤200ms







                  GR


                  Output current

                  Range

                  3.00A~52.00A(DC)

                  Accuracy

                  ±(1.5% Value +0.2A)

                  Resolution

                  0.01A

                  Maximum output power

                  250W(50A/0.10Ω)

                  Maximum output current

                  50AW

                  Resistance upper limit setting value

                  40A x150mW

                  Set current value

                  Output waveform

                  Sine wave

                  Output waveform distortion

                  ≤2% (no load or pure resistive load)

                  Crest factor

                  1.3~1.5

                  Output signal type

                  DDS+power amplifier

                  Test time

                  0.3s~999.9s 0=continuous test

                  Interval time

                  0.3s~999.9s 0=interval time off

                  Voltmeter


                  Range

                  0.050kV~6.500kV

                  0.050kV~6.500kV

                  0.050kV~6.500kV

                  0.050kV~6.500kV

                  Accuracy

                  ±(1.5% reading +5V)

                  Resolution

                  1V

                  Display value

                  RMS value



                  Withstand voltage ammeter

                  Measure range

                  AC

                  0 ~ 20mA

                  DC

                  0 ~ 10mA


                  Resolution

                  AC

                  200uA: 0.1uA, 2mA: 1uA, 20mA: 10uA

                  DC

                  2uA: 0.001uA, 20uA: 0.01uA, 200uA: 0.1uA, 2mA: 1uA,

                  10mA: 10uA

                  Measurement accuracy

                  ≥2mA: ±(1.5%+5 digits), <2mA: ±(2%+5 digits)

                  Offset function

                  The current of the test leads and accessories can be subtracted.

                  Test mode

                  GND mode: RETURN terminal connected to the chassis

                  FLOAT mode: RETURN terminal not connected to the chassis


                  Measurement range

                  1MΩ~50000MΩ


                  Resolution

                  1MΩ~9.999MΩ:0.001MΩ,10MΩ~99.99MΩ:0.01MΩ,

                  100MΩ~999.9MΩ:0.1MΩ,1000MΩ~9999MΩ:1MΩ

                  10GΩ~50GΩ:10MΩ

                  Insulation resistance meter



                  Accuracy

                  0.100kV~0.200kV:1MΩ~999MΩ ±5%,1000MΩ~3000MΩ ±10%

                  0.201kV~0.499kV:1MΩ~999MΩ ±5%,1000MΩ~5000MΩ ±10%

                  0.500kV~1.000kV:1MΩ~999MΩ ±5%,1000MΩ~9999MΩ ±10%

                  10000MΩ~50000MΩ ±15%

                  Ground current meter

                  Measuring range

                  3.00A~52.00A

                  Resolution

                  0.01A

                  Accuracy

                  ±(1.5% of reading + 0.2A)

                  Display value

                  RMS value

                  Ground resistance meter

                  Range

                  0~510mΩ

                  Accuracy

                  When the current is less than 5.00A, it is ± (3% of reading + 3mΩ) When the current is ≥ 5.00A, it is ± (2% of reading + 2mΩ)

                  Resolution

                  0.1mΩ

                  Measuring method

                  Four-terminal method

                  Timer

                  Range

                  0~999.9s

                  Resolution

                  0.1s

                  Accuracy

                  ±(0.1%+50ms)

                  MEcjcjavascript:void();