• <pre id="HEMiqQ"></pre>

          1. <legend id="HEMiqQ"><tr id="HEMiqQ"></tr></legend>

                  Proof Tracking Index tester / Proof Tracking Test Apparatus/Dielectric Tester

                  The proof Tracking tester for electric and electronic products, household appliances, and its material is resistant to leakage from the mark test, to simulate the live parts of different polarity or between live parts and grounded metal parts may cause the insulator leakage measurement and evaluation of the specified voltage, the insulator in the relative resistance to leakage when the water effect of the electric field and impurities from the mark.
                  Description

                  Product Description:

                  Name:Proof Tracking Index tester / Proof Tracking Test Apparatus/Dielectric Tester

                  Type:BND-LD


                  Product Description

                  IEC60335-1Proof Tracking Index Tester for Leakage Current test

                  The proof Tracking tester for electric and electronic products, household appliances, and its material is resistant to leakage from the mark test, to simulate the live parts of different polarity or between live parts and grounded metal parts may cause the insulator leakage measurement and evaluation of the specified voltage, the insulator in the relative resistance to leakage when the water effect of the electric field and impurities from the mark. The device in accordance with national standards GB/T4207-2003 (IEC60112-2003) "solid insulating materials under moist conditions compared to the determination method of tracking index and index of resistance to tracking and GB4706.1-2005" Household and similar electrical appliances - Safety part I: General Requirements set forth in the relevant provisions of the design and manufacture.


                  Detailed Images

                  the electrode materialrectangular electrode - platinum metal
                  the electrode pressure1.0 ± 0.05N
                  the distance between electrodes4.0mm ± 0.01mm, angle 60 ° ± 5
                  the electrode voltage100 ~ 600V (48 ~ 60HZ) adjustable short-circuit current 1.0 ± 0.1A, voltage drop is less than 10%
                  drops of liquid height30 ~ 40mm adjustable
                  the size of the drops44 to 55 drops / 1cm3 Adjustable
                  drops Time30s ± 5s,
                  the number of drops0 to 9999 times adjustable
                  the leakage judgeloop current greater than 0.5A and maintain the 2 seconds the relay to cut off the current, indicating that the sample failed
                  the test processautomatic control of test procedures, independent ventilation
                  reference standardsIEC60112-2003, GB/T4207-2003, GB4706.1-2005
                  the studio volume0.5 cubic
                  the device outer dimensions1000mm wide x 650mm deep x 1300mm high


                  Packing & Delivery